CAI offers four analyzer models that utilize Fourier Transfer Infrared technology for measuring concentrations of multiple gases within a sample. CAI FTIR analyzers deliver fast, accurate analysis of virtually any gas that has an infrared absorption spectrum.
- 7000 FTIR Heated Sampler – Most IR-absorbing gases + short cell for higher concentration levels
CAI FTIR analyzers are based on Fourier Transform Infrared Spectroscopy. Non-symmetrical gas phase molecules absorb IR light, causing the molecular bonds to stretch, bend or rotate. This absorption is used to measure and quantify several chemical components simultaneously.
- Brand
- Overview
- Features
- Product Data Sheet
Overview
The 7000 Series FTIR is based on Fourier Transform Infra-red Spectroscopy. Nonsymmetrical gas phase molecules absorb IR light which in turn causes the molecular bonds to stretch, bend, or rotate. This absorption is used to measure and quantify several chemical components simultaneously. An IR source emits radiation in the range of 7500 to 375 cm-1 . The IR radiation is
split in a Michelson Interferometer where half of the light passes through to a fixed mirror and the other half is reflected
towards a moving mirror. The two beams recombine and pass through a gas cell where the sample absorbs light at molecule specific frequencies. The remaining light is measured with a DTGS detector and Fourier transformed to convert from time domain to frequency domain producing a single beam spectrum which is ratioed with a baseline spectrum and produces an absorbance spectrum. This absorbance spectrum is quantified with chemometrics to produce a concentration value.
Features
NEW Touch screen display provides temperature control circuits for oven, pump, and heated sample line to analyzer (5 feet)
Ensures critical components are maintained at 191 o C
Single –head heated pump, 5.5 LPM (Optional HI-flow with Variable 11-14 LPM)
T-Handle heated filter (S21-T type borosilicate element, 0.01 micron) located on front panel for easy access
Heavy duty high-temperature flow meter
Solenoid valves for zero, span and bias calibrations
Electrically actuated ball valve for switching between sample and calibration modes
Stainless-steel tubing and 1/4” SS bulkhead connection fittings
Modbus TCP I/O for control and outputs
Options
700 FTIR Temperature Controller Accessory provides up to 4 circuits for external heated lines and/or heated-jacket
17” W x 8” D x 3.5” H)
Digital inputs to control valves
Power requirements: 115 VAC/9.3 amps; 230 VAC/4.78 amps
19-inch rack mountable, fan-cooled enclosure (17” W x 22” D x 8.5” H)