YSI’s new IQ SensorNet Analyzer for Orthophosphate provides accurate, reliable measurements for process control. The IQ Analyzer features a revolutionary MultiValve that dramatically reduces reagent use and maintenance requirements.
- Brand
- OVERVIEW
- SPECIFICATIONS
OVERVIEW
The Alyza IQ PO4 is an online, cabinet-style analyzer used for continuous monitoring of orthophosphate in a municipal water resource recovery facility. This analyzer is weatherproof and can be deployed anywhere in the wastewater treatment process. Applications include monitoring bio-p removal and monitoring and controlling chemical phosphorus removal. An innovative Multivalve within the analyzer significantly reduces reagent use to 5 µL per measurement decreasing the analyzer’s regular maintenance requirements. The analyzer is available as a single-channel or two-channel instrument. The analyzer is available in one of two measuring ranges: 0.02 to 15.00 mg/L PO₄-P or 0.2 to 50.0 mg/L PO₄-P. The IQ Analyzer can be used as a stand-alone analyzer or be incorporated into an IQ SensorNet network of sensors. The IQ Analyzer provides accurate measurements with minimal maintenance requirements.
IQ Analyzer offers:
- Reliable, accurate measurements at low measurement concentrations
- Extremely low reagent consumption, 5 µL per measurement
- Automatic calibration, 1 or 2 point
- Automatic cleaning, lowering maintenance requirements
- Serviceable – safe and easy replacement of reagents, no need for expensive service contracts
- One or two channel versions available
- Reliable sample delivery pump built into the cabinet
- Uses proven Vanadomolybdate (yellow) method of detection
- Filtration unit allows analyzer to process samples with solids concentrations up to 6,000 mg/L
Applications include:
- Monitoring and control of chemical phosphorus removal, reducing dosing chemicals and sludge production
- Effluent monitoring to ensure compliance
SPECIFICATIONS
| Calibration | Automatic 1- and 2-point calibrations |
|---|---|
| Cleaning | Automatic cleaning with cleaning solutions |
| Connectivity / Communications | IQ Sensor Net-components, relays, Modbus TCP, analog outputs |
| Dimensions | 30.5 x 26.6 x 17.5 in. |
| Dissolved Solids Limit | < 6 g/L (before filtration) |
| Equipment used with | IQ SensorNet 2020 3G Controller, 282/284 Controller |
| Languages | English, German |
| Measurement Method | Vanadomolybdate (yellow) method |
| Operating Temperature | -4 to 104 °F (-20 to +40) °C |
| Parameters Measured | Orthophosphate |
| Power | 120 VAC / 240 VAC, 50/60 Hz |
| Reagent Use | 5 µL per measurement |
| Sample Streams/Channels | 1- and 2-channel versions available |
| Sample Temperature | 39 to 113 °F (4 to 45 °C) |
| Warranty | 2 Years |
| Weight | 81.6 lbs (37 kg) (without liquids) |
| PO4-111/112 | PO4-121/122 | |
|---|---|---|
| Measuring Range |
(Instrument Dependent)MR 1: 0.02 to 15.00 mg/l PO4-PMR 2: 0.2 to 50.0 mg/l PO4-PDisplayed0.00 to 15.00 mg/l PO4-P0.0 to 50.0 mg/l PO4-PResolution0.01 mg/l PO4-P0.05 mg/l PO4-PAccuracy± 2 %, ± 0.02 mg/l± 2 %, ± 0.2 mg/l