Model 3010TBC Oxygen Analyzer

SKU: 3010TBC

​The Model 3010C Series BASEEFA/CENELEC approved, intrinsically safe trace oxygen analyzers are versatile, microprocessor based instruments used for detecting oxygen from 0-10 ppm to 100% in a variety of gases.

  • Brand
  • DETAILS
  • FEATURES
  • SPECIFICATION
  • APPLICATIONS

Brand

Teledyne Analytical TAI

DETAILS

Maintenance-Free

The series features a Micro-fuel Cell to measure oxygen in the 0-10 ppm to 100% range. The sensors set industry standards for accuracy, sensitivity and ease of use, and come with a one year limited warranty against defects in material and workmanship. Also, because every sensor is manufactured using stringent quality control procedures, the user is assured of outstanding reliability and optimum performance. The sensor is a sealed electrochemical device with no electrolyte to change or electrodes to lean, making it maintenance free. The sensor is specific to oxygen and can be used in a variety of background gases including hydrocarbons.

Flexibility

These units offer three dynamic, user-configurable ranges and can be calibrated at any range. Two fully programmable concentration alarms and relays provide the versatility to satisfy nearly any requirement.

Convenience

Two standard 0-1 VDC outputs and two 4-20mADC isolated outputs provide both concentration and range identification. A bi-directional RS-232 serial interface is incorporated to relay information to a host computer for remote monitoring of zero and span calibration.

Split Architecture

These units are designed as “split architecture” instruments, meaning that a general purpose control unit for non-hazardous areas controls a specially designed analysis unit or “remote probe” that is CENELEC approved intrinsically safe for use in area classification EEXiallCT6, Zone 0. Simple menu choices, membrane command switches and dual displays make setup and operation clear and quick. The 3010TC series is constructed of high quality materials to insure long term, trouble free materials to insure long term, trouble free service.

Options

  • K        19″ rack mount with either one or two units installed, mounted side by side (flush mount versions only)
  • S        Stainless steel cell holder

Alternate Sensors

Class L2C-L: Backgrounds with CO2 90% of FS in < 60 seconds; up to 1 year life

FEATURES

  • Maintenance-free Micro-Fuel Cell
  • Three user-selectable ranges plus cal range
  • Signal output: 0-1 VDC & 4-20 mADC
  • Programmable auto ranging
  • Range ID contacts (qty 4), Form A normally open contacts, 3A resistive
  • Two fully-adjustable concentration alarm points with programmable relay function, Form C contacts, 3A resistive
  • Programmable auto cal/zero with Form A normally open contact relay signals
  • Remotely indicated cal/zero via customer supplied 24 VDC signal and 12 VDC control valves
  • Self diagnostics with Form C failure alarm contacts
  • Full duplex RS-232 communication link
  • Five digit oxygen concentration LED display
  • Back-lit 2 x 20 line alphanumeric vacuum fluorescent display for setup and diagnostics
  • Universal power supply: 85-230 VAC, 50/60 Hz
  • Remote probe, nylon with stainless steel fittings
  • Oxygen sensor:
  • ​3010TC    Micro-Fuel Cell, Class L2C, up to 2 years life
  • 3010PC    Micro-Fuel Cell, Class C-3,  up to 1.5 years life

SPECIFICATION

​Ranges​ ​3 customer selectable ranges
3010TC                          Min: 0-10 ppm
Max: 0-25%
3010PC                          Min: 0-1%
Max: 0-100%
Calibration Range:      0-25% (unit can be calibrated on any range or gas)
​Accuracy ​± 2% of FS at a constant temperature
± 5% of FS over operating temperature range
(once temperature equilibrium has been reached; except 0-10 ppm scale which is ± 1 ppm)
​Sensitivity ​0.5% of FS
​Response Time 3010TC:         < 60 seconds
3010PC:         < 30 seconds
​Temperature ​32° to 122°F (0° to 50°C)
​Output ​Analytical:     0-1 VDC and 4-20 mADC isolated, 1000 ohm impedance
Range ID:      0-1 VDC and 4-20 mADC isolated
​Power ​83 to 230 VAC, 50/60 Hz
​Enclosure ​3010TAC:      General purpose, flush panel mount
3010TBC:      I.P. 65/66 rated wall mount
​Dimensions ​3010TAC & PAC:       Case:     8.7″W x 6.96″H x 12.2″D
(20.1 x 17.68 x 30.99 cm)​
Bezel:   10.79″W x 7.46″H
(27.4 x 18.95 cm)
3010TBC & PBC:                     25.63″H x 15.25″W x 11″D
(65.1 x 38.74 x 27.94 cm)​

APPLICATIONS

  • Air separation and liquefaction
  • Pure gaseous hydrocarbon stream monitoring
  • Semiconductor manufacturing
  • Protective atmosphere blanketing of primary liquid feedstocks​​ and flammable liquids
  • Process monitoring of gaseous monomers, vinyl chloride, propylene, butadiene, isoprene or ethylene
  • Gas purity certification
More From Avensys

More Services

Our certified engineers and technicians can intervene from the initial consultation and planning stages, throughout the life of instruments and systems.
Happy Professional Heavy Industry Engineer Worker Wearing Uniform, and Hard Hat in a Steel Factory. Smiling Industrial Specialist Standing in a Metal Construction Manufacture.

Field & In House

Develop customized monitoring plans that address client's unique needs...
https://avensys.com/wp-content/uploads/2023/05/power-plant-1280x854.jpg

Start up

Start up & Commissioning is a critical phase in the life cycle of any project.
https://avensys.com/wp-content/uploads/2023/05/integrated2500-1280x849.jpg

Integrated Systems

The benefits include improved performance, reduced costs, and increased reliability.
https://avensys.com/wp-content/uploads/2023/05/Depositphotos_445297446_XL-1280x854.jpg

Training

We believe that product training is essential for both our employees and our clients.