| Optics | temperaturecompensated LED ( λ = 860 nm) and photosensor amplifier in water proof sample chamber, infrared light |
| Measuring Principle | nephelometric (90° scattered light) |
| Measuring Range | 0,01-1100 NTU (Autorange) |
| Accuracy | ± 2.5 % of reading, or ± 0.01 NTU, whichever is bigger, in the range of 0.01 – 500 NTU ; ± 5 % of reading in the range of 500 – 1000 NTU |
| Display | backlit LCD |
| Interfaces | Micro-USB |
| Operation | polycarbonate membrane, splash proof |
| Auto – OFF | Yes |
| Internal Storage | internal ring memory for 125 data sets |
| Power Supply | 9 V-block |
| Clock | Real Time Clock and Date |
| Portability | Portable |
| Environmental Conditions | 5 – 40 °C at 30 – 90 % relative humidity (non condensing) |
| Compliance | CE |
| Dimensions | 110 x 55 x 190 mm |
| Weight | 400 g (Basic unit) |
| Weight with Packaging | (Basic unit) |
| Testparameter | Measuring Range | Method No. |
|---|---|---|
| Turbidity | 0.01 – 1100 NTU |